Tube Scoop:
Insider's Guide to Cleaning Lap Instruments
Insulation Inspection
Among reusable laparoscopic instruments, there are
basically two types: insulated and non insulated. The first thing to know about the
insulation on laparoscopic instruments is that it's temporary. It's got to be
monitored for nicks and cuts, and many times it's got to be replaced. Checking that
insulation is part of your job in central service.
There are safeguards tin the operating room to reduce
electrical risks. But when the insulation gets scratched of nicked so that metal
shows through, the electricity that's transmitted to the distal end of an instrument can
arc to the neighboring tissue or organs during a procedure and cause surgical
complications. And there are risks to surgeons as well. There are reports that
insulation failures in an instrument's handle rings can enable electrical charges to reach
a surgeon's glove.
Teach your staff how to inspect for flaws in the
insulation. The technique is to begin by inspecting the distal tip. Look at
the first three inches very closely for insulation failure. Those few inches are
where you'll most often find insulation failure.
When inspecting insulation, you need a 360-degree view of
the instrument. Start at the distal tip and slowly rotate the instrument, examining
it all the way down to the rings. The rings are another place where you may find
cracks in the insulation. you may even find that the insulation is completely
missing.
When you discover insulation failures, communicate your
findings to the appropriate clinical authority. Such communication is a way to
market your department as a center of quality control.
Nicks and scratches aren't the only problems you have to
monitor. Many times the insulation, which manufacturers heat-shrink onto an
instrument, can loosen and pull away from the distal end. Make sure the insulation
hasn't become loose and able to slide. For quality assurance, hold on to the shaft
of the instrument and pull back on the insulation.
If it slides, schedule that instrument to be repaired.
Sliding insulation allows more metal to be exposed than should be. Another negative
effect: insulation that slides can conceal surgical debris. Once again, communicate
such discoveries to your clinical customers in the surgery department.
Shaft reinsulation is not a complicated repair and
shouldn't take more than 24 hours. The cost: $25 to $60. Handle reinsulation,
however, is a more complicated repair. It can take seven to ten days and cost $100
to $250.
Reinsulation is just one of the many types of repairs that
can be done to laparoscopic instruments. When a vendor tells you that you have an
instrument that can't be repaired, always get a second opinion. Usually, the only
cost involved in getting a second opinion is that of shipping, which can be as low as $2
to $4. Savings on a misdiagnosed instrument can easily be $400 to $600. It can
be tedious to find a high-quality repair vendor in your market to provide a second
opinion, but the savings can be tremendous.
[ Next ]